* For a detailed description of our capabilities, ranges and accuracy’s, refer to our current accredited schedule of pressure calibration, please click here.
Pressure Range
Gauge and Absolute mode in the pressure range -0.1 MPa to +86 MPa pneumatic and +0.5 MPa to +140 MPa hydraulic.
Dead-weight Testing Methods
#1 Pressure Equivalent - The certificate of calibration shows the pressure generated by certain combinations of weights applied to the dead-weight tester.
#2 Effective Area - The certificate of calibration shows information that, when combined with mass and local gravity values, can be used to:
- Accurately calculate the pressure generated by any combination of weights applied to the dead-weight tester.
- Accurately calculate the weight combination that must be applied to the dead-weight tester to generate the required pressure.
|